These autoclaves are
specially designed for biological safety laboratories and clean rooms.

Safely linking
different areas within a laboratory, the double-door autoclaves are built into the wall between two rooms and provide a completely gas tight interface.
Removing the need to transport items to and from the autoclaving area using trolleys, this autoclave allows controlled and direct transfer of sterilised
items into the laboratory, and of contaminated items out of the laboratory for sterilisation and safe disposal. Only one door can be opened at a time as both
doors are interlocked.
Features
The 2D autoclaves are available in 95, 150 and 200 litre volumes,
and are part of a range that brings together reduced cycle times, innovative design and intelligent software to meet the needs of today’s laboratory.
Advanced system and process technology has been used to optimise temperature and pressure regulation within the chamber, and increase the rate of the heating
phase by up to 50% when compared with conventional autoclaves. Spray cooling is one of a number of options available to significantly reduce cooling times,
by up to 90%.
Exhaust steam is condensed automatically, which prevents odours and protects any plastic
waste water piping.
Triple safety
features
For additional safety, the 2D series has three special safety features:
- only one door can be opened at a time as both doors are interlocked
- if the autoclave is switched off or there is a loss of power, both doors will remain locked
- if
the door in the non-sterile environment is opened, a sterilisation programme has to be performed before the door in the sterile room can be opened.
Operation can, however, be carried out from both sides of the autoclave as the display indicates whether the
opposite door is open or closed.
Calibration and
Validation
To ensure regulatory compliance, all models in the range can be fully validated and
offer a choice of 12 sterilisation programmes and an optional printer to provide detailed records of all essential parameters. Validation information can
also be downloaded to a PC, for which easy to use software is available for comprehensive data manipulation and diagnostic protocols.
With UKAS laboratory accreditation and BS EN ISO 9001:2000 approval, Don Whitley Scientific can provide calibration and
validation services for all autoclaves.
Pressure safe up
to 5 bar/150°C
The Systec D-Series 2D is the first autoclave of its kind to be designed for higher pressures
and temperatures. The standard pressure chamber is designed to cope with 5 bar/150°C. They are also equipped for sterilisation temperatures up to
140°C and a pressure of 4 bar.
|